Specifications

  • Operating temperature: 15-35 °C
  • Storage temperature: 5-45 °C
  • Power supply 230 V (external power supply unit 12V 1.5A)
  • Compatible syringes: 10, 20, 50 ml
  • Adjustable dosage from 1 to 300 ml/h in steps of +/- 0.1 ml/h
  • Bolus function
  • Slide travel: 170 mm
  • Force sensor: analogue 0.15 to 20 N
  • Position sensor: 0.6 mm resolution
  • End of injection alarm adjustable from 0 to 120 minutes or from 0 to 10 ml before the end of injection
  • LCD display and simple user interface (3 push buttons + 1 dedicated to security)
  • Indication of the remaining injection time, the quantity already administered and the percentage of syringe filling
  • Possibility of chaining three EPS supplied with 12V DC by a single mains supply sized accordingly.
  • Protection of the external 12V DC power supply against short-circuits: by an internal fuse with automatic reset.
  • Mounting on hospital bed or mobile vertical column